USGS · ofr94358
Inductively coupled plasma-mass spectrometric method for the determination of dissolved trace elements in natural water
Abstract
An inductively coupled plasma-mass spectrometry method was developed for the determination of dissolved Al, As, B, Ba, Be, Cd, Co, Cr, Cu, Li, Mn, Mo, Ni, Pb, Sr, Tl, U, V, and Zn in natural waters. Detection limits are generally in the 50-100 picogram per milliliter (pg/mL) range, with the exception of As which is in the 1 microgram per liter (ug/L) range. Interferences associated with spectral overlap from concomitant isotopes or molecular ions and sample matrix composition have been identified. Procedures for interference correction and reduction related to isotope selection, instrumental operating conditions, and mathematical data processing techniques are described. Internal standards are used to minimize instrumental drift. The average analytical precision attainable for 5 times the detection limit is about 16 percent. The accuracy of the method was tested using a series of U.S. Geological Survey Standard Reference Water Standards (SWRS), National Research Council Canada Riverine Water Standard, and National Institute of Standards and Technology (NIST) Trace Elements in Water Standards. Average accuracies range from 90 to 110 percent of the published mean values.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J.R. Garbarino, Howard E. Taylor. 1996. Inductively coupled plasma-mass spectrometric method for the determination of dissolved trace elements in natural water. https://doi.org/10.3133/ofr94358
Cite the original work for its findings. Save a collection to share your selection of sources.