Reduction of matrix interferences in furnace atomic absorption with the L'vov Platform

Abstract

Use of a modified L'vov Platform and ammonium phosphate as a matrix modifier greatly reduced matrix interferences in a commercial Massmann-type atomic absorption furnace. Platforms were readily fabricated from furnace tubes and, once positioned in the furnace, caused no inconvenience in operation. Two volatile elements (Pb, Cd), two of intermediate volatility (Co, Cr) and two which form stable oxides (Al, Sn) were tested in natural water and selected synthetic matrices. In every case for which there was a significant matrix effect during atomization from the tube wall, the platform and platform plus modifier gave improved performance. With lead, for example, an average ratio of 0.48 ± 0.11 was found when the slope of the standard additions plot for six different natural water samples was compared to the slope of the standard working curve in dilute acid. The average slope ratio between the natural water matrices and the dilute acid matrix was 0.94 ± 0.03 with the L'vov Platform and 0.96 ± 0.03 with the platform and matrix modifier. In none of the cases studied did the use of the platform or platform plus modifier cause an interference problem where none existed while atomizing from the tube wall. An additional benefit of the platform was a factor of about two improvement in peak height precision.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M. L. Kaiser, S. R. Koirtyohann, E. J. Hinderberger, Howard E. Taylor. 1981. Reduction of matrix interferences in furnace atomic absorption with the L'vov Platform. https://doi.org/10.1016/0584-8547(81)80060-2

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related USGS reports

Improving ChemCam LIBS long-distance elemental compositions using empirical abundance trends

The ChemCam instrument on the Curiosity rover provides chemical compositions of Martian rocks and soils using remote laser-induced breakdown spectroscopy (LIBS). The elemental calibration is stable as a function of distance for Ti, Fe, Mg, and Ca. The calibration shows small, systematically increasing abundance trends as a function of distance for Al, Na, K, and to some extent, Si. The distance effect is known to be due to a dependence with distance on the relative strengths of atomic transition lines. Emission lines representing transitions from relatively low energy levels remain intense at longer distances while emission lines representing transitions from higher energy levels decrease in intensity more rapidly as a function of distance. The multivariate algorithms used to determine elemental compositions rely on a large number of emission lines in many cases, so rather than trying to correct all emission lines, a study was made of the predicted compositions as a function of distance, in order to determine an empirical correction. Abundance trends can be well approximated by a linear trend with distance within the ranges of abundances and distances observed up to ~6 m. Data from 11 distinct geological members and data groups of the Murray formation in Gale crater, Mars, were used to form the model, selecting the members and data groups yielding the best statistics. The model was tested using data from several targets observed from two different distances, and using data from the Kimberley formation, the composition of which is significantly different from the Murray formation, showing that the model works on other compositions beyond those used to build the model. For long-distance observations up to ~6 m, corrections can be made back to an equivalent composition at the median distance of ChemCam observations (2.6 m). The model has been validated up to 6.2 m, although ChemCam is able to observe bedrock targets to >7 m, and iron meteorites to distances of >9 m.

Spectrochimica Acta Part B: Atomic Spectroscopy

Improved accuracy in quantitative laser-induced breakdown spectroscopy using sub-models

Accurate quantitative analysis of diverse geologic materials is one of the primary challenges faced by the Laser-Induced Breakdown Spectroscopy (LIBS)-based ChemCam instrument on the Mars Science Laboratory (MSL) rover. The SuperCam instrument on the Mars 2020 rover, as well as other LIBS instruments developed for geochemical analysis on Earth or other planets, will face the same challenge. Consequently, part of the ChemCam science team has focused on the development of improved multivariate analysis calibrations methods. Developing a single regression model capable of accurately determining the composition of very different target materials is difficult because the response of an element’s emission lines in LIBS spectra can vary with the concentration of other elements. We demonstrate a conceptually simple “sub-model” method for improving the accuracy of quantitative LIBS analysis of diverse target materials. The method is based on training several regression models on sets of targets with limited composition ranges and then “blending” these “sub-models” into a single final result. Tests of the sub-model method show improvement in test set root mean squared error of prediction (RMSEP) for almost all cases. The sub-model method, using partial least squares regression (PLS), is being used as part of the current ChemCam quantitative calibration, but the sub-model method is applicable to any multivariate regression method and may yield similar improvements.

Spectrochimica Acta Part B: Atomic Spectroscopy

Recalibration of the Mars Science Laboratory ChemCam instrument with an expanded geochemical database

The ChemCam Laser-Induced Breakdown Spectroscopy (LIBS) instrument onboard the Mars Science Laboratory (MSL) rover Curiosity has obtained > 300,000 spectra of rock and soil analysis targets since landing at Gale Crater in 2012, and the spectra represent perhaps the largest publicly-available LIBS datasets. The compositions of the major elements, reported as oxides (SiO 2 , TiO 2 , Al 2 O 3 , FeO T , MgO, CaO, Na 2 O, K 2 O), have been re-calibrated using a laboratory LIBS instrument, Mars-like atmospheric conditions, and a much larger set of standards (408) that span a wider compositional range than previously employed. The new calibration uses a combination of partial least squares (PLS1) and Independent Component Analysis (ICA) algorithms, together with a calibration transfer matrix to minimize differences between the conditions under which the standards were analyzed in the laboratory and the conditions on Mars. While the previous model provided good results in the compositional range near the average Mars surface composition, the new model fits the extreme compositions far better. Examples are given for plagioclase feldspars, where silicon was significantly over-estimated by the previous model, and for calcium-sulfate veins, where silicon compositions near zero were inaccurate. The uncertainties of major element abundances are described as a function of the abundances, and are overall significantly lower than the previous model, enabling important new geochemical interpretations of the data.

Spectrochimica Acta Part B: Atomic Spectroscopy