Abstract
A method for determination of cobalt-60 in waters at levels greater than 0.5 pCi per sample is presented. A modification of the method may be used to analyze fluvial sediments and soils. After the cobalt has been separated, first as the hydroxide and then as the thiocyanate complex in methyl isobutyl ketone, it is counted in a liquid scintillation system at 80% efficiency. Separation factors achieved for six isotopes are generally greater than 2,000. The time for a single analysis, exclusive of the counting and evaporation operations, is about 2 h.