USGS · 70004044
Interlaboratory comparison of measurements of acid-volatile sulfide and simultaneously extracted nickel in spiked sediments
Abstract
An interlaboratory comparison of acid-volatile sulfide (AVS) and simultaneously extracted nickel (SEM_Ni) measurements of sediments was conducted among five independent laboratories. Relative standard deviations for the seven test samples ranged from 5.6 to 71% (mean = 25%) for AVS and from 5.5 to 15% (mean = 10%) for SEM_Ni. These results are in stark contrast to a recently published study that indicated AVS and SEM analyses were highly variable among laboratories.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
William G. Brumbaugh, Chad R. Hammerschmidt, Luciana Zanella, Emily Rogevich, Gregory Salata, Radoslaw Puchalka. 2011-06-01. Interlaboratory comparison of measurements of acid-volatile sulfide and simultaneously extracted nickel in spiked sediments. https://doi.org/10.1002/etc.506