Recovery of several volatile organic compounds from simulated water samples: Effect of transport and storage
No abstract available.
Geology topics
Publications and source records attributed to L.J. Schroder.
No abstract available.
Since the inception of the U.S. Geological Survey National Field Quality Assurance Program, over 85,000 proficiency samples have been analyzed by water-quality analysts. This includes more than 10,000 alkalinity samples, more than 15,000 pH samples, and more than 16,000 specific conductance samples, which were analyzed from 1990 through 1997, and a total of more than 43,000 proficiency samples analyzed from 1979 through 1989. The analyte values were evaluated to determine the fourth-spread, a measure of the width of the middle half of the data, and the F-pseudosigma, a robust replacement for the standard deviation, for each of the different measurement ranges. The result of the statistical evaluation showed that the vast majority of reference sample measurements made by water-quality analysts were within acceptable ranges. From 1990 to 1997, the measurement of pH had the highest level of acceptable results, 98.4 percent, followed by specific conductance with 95.2 percent acceptable results, and alkalinity with 88.6 percent acceptable results. The statistical summary of pH indicates the calculated fourth-spread values for the entire tested range is +0.06 pH units. For specific conductance, the magnitude of the fourth-spread increases as the magnitude of the specific conductance ranges increases. The average relative fourth-spread percent for all reported specific conductance values is +1.8 percent. From 1990 through 1997, the evaluation of the results for alkalinity measurement for the average fourth-spread was determined to be + 3.3 milligrams per liter as calcium carbonate.
The U.S. Geological Survey used four programs in 1991 to provide external quality assurance for the National Atmospheric Deposition Program/National Trends Network (NADP/NTN). An intersite-comparison program was used to evaluate onsite pH and specific-conductance determinations. The effects of routine sample handling, processing, and shipping of wet-deposition samples on analyte determinations and an estimated precision of analyte values and concentrations were evaluated in the blind-audit program. Differences between analytical results and an estimate of the analytical precision of four laboratories routinely measuring wet deposition were determined by an interlaboratory-comparison program. Overall precision estimates for the precipitation-monitoring system were determined for selected sites by a collocated-sampler program. Results of the intersite-comparison program indicated that 93 and 86 percent of the site operators met the NADP/NTN accuracy goal for pH determinations during the two intersite-comparison studies completed during 1991. The results also indicated that 96 and 97 percent of the site operators met the NADP/NTN accuracy goal for specific-conductance determinations during the two 1991 studies. The effects of routine sample handling, processing, and shipping, determined in the blind-audit program indicated significant positive bias (a=.O 1) for calcium, magnesium, sodium, potassium, chloride, nitrate, and sulfate. Significant negative bias (or=.01) was determined for hydrogen ion and specific conductance. Only ammonium determinations were not biased. A Kruskal-Wallis test indicated that there were no significant (*3t=.01) differences in analytical results from the four laboratories participating in the interlaboratory-comparison program. Results from the collocated-sampler program indicated the median relative error for cation concentration and deposition exceeded eight percent at most sites, whereas the median relative error for sample volume, sulfate, and nitrate concentration at all sites was less than four percent. The median relative error for hydrogen ion concentration and deposition ranged from 4.6 to 18.3 percent at the four sites and as indicated in previous years of the study, was inversely proportional to the acidity of the precipitation at a given site. Overall, collocated-sampling error typically was five times that of laboratory error estimates for most analytes.
A collocated, wet-deposition sampler program has been operated since October 1988 by the U.S. Geological Survey to estimate the overall sampling precision of wet atmospheric deposition data collected at selected sites in the National Atmospheric Deposition Program and National Trends Network (NADP/NTN). A duplicate set of wet-deposition sampling instruments was installed adjacent to existing sampling instruments at four different NADP/NTN sites for each year of the study. Wet-deposition samples from collocated sites were collected and analysed using standard NADP/NTN procedures. Laboratory analyses included determinations of pH, specific conductance, and concentrations of major cations and anions. The estimates of precision included all variability in the data-collection system, from the point of sample collection through storage in the NADP/NTN database. Sampling precision was determined from the absolute value of differences in the analytical results for the paired samples in terms of median relative and absolute difference. The median relative difference for Mg2+, Na+, K+ and NH4+ concentration and deposition was quite variable between sites and exceeded 10% at most sites. Relative error for analytes whose concentrations typically approached laboratory method detection limits were greater than for analytes that did not typically approach detection limits. The median relative difference for SO42- and NO3- concentration, specific conductance, and sample volume at all sites was less than 7%. Precision for H+ concentration and deposition ranged from less than 10% at sites with typically high levels of H+ concentration to greater than 30% at sites with low H+ concentration. Median difference for analyte concentration and deposition was typically 1.5-2-times greater for samples collected during the winter than during other seasons at two northern sites. Likewise, the median relative difference in sample volume for winter samples was more than double the annual median relative difference at the two northern sites. Bias accounted for less than 25% of the collocated variability in analyte concentration and deposition from weekly collocated precipitation samples at most sites.A collocated, wet-deposition sampler program has been operated since OCtober 1988 by the U.S Geological Survey to estimate the overall sampling precision of wet atmospheric deposition data collected at selected sites in the National Atmospheric Deposition Program and National Trends Network (NADP/NTN). A duplicate set of wet-deposition sampling instruments was installed adjacent to existing sampling instruments four different NADP/NTN sites for each year of the study. Wet-deposition samples from collocated sites were collected and analysed using standard NADP/NTN procedures. Laboratory analyses included determinations of pH, specific conductance, and concentrations of major cations and anions. The estimates of precision included all variability in the data-collection system, from the point of sample collection through storage in the NADP/NTN database.
The U.S. Geological Survey has a policy that requires each District office to prepare a Quality Assurance Plan. This plan is a combination of a District's management principles and quality assurance processes. The guidelines presented in this report provide a framework or expanded outline that a District can use to prepare a plan. Parti- cular emphasis is given to a District's: (1) quality assurance policies; (2) organization and staff responsibilities; and (3) program and project planning. The guidelines address the 'how', 'what', and 'who' questions that need to be answered when a District Quality Assurance Plan is prepared.
Explore the source record for details and available documents.
Explore the source record for details and available documents.
Explore the source record for details and available documents.
A continuing quality-assurance program has been operated by the U.S. Geological Survey to evaluate any bias introduced by routine handling, shipping, and laboratory analyses of wet-deposition samples collected in the National Atmospheric Deposition Program (NADP) and National Trends Network (NTN). Blind-audit samples having a variety of constituent concentrations and values were selected. Only blind-audit samples with constituent concentrations and values less than the 95th-percentile concentration for natural wet-deposition samples were included in the analysis. Of the major ions, there was a significant increase of Ca 2+ , Mg 2+ , Na 2+ , K + , SO 4 2− and Cl −1 in samples handled according to standard protocols and shipped in NADP/NTN sample-collection buckets. For 1979–1987, graphs of smoothed data showing the estimated contamination in blind-audit samples indicate a decrease in the median concentration and ranges of Ca 2+ , Mg 2+ and SO 4 2− contamination of blind-audit samples shipped in sample-collection buckets. Part of the contamination detected in blind-audit samples can be attributed to contact with the sample-collection bucket and lid; however, additional sources also seem to contaminate the blind-audit sample. Apparent decreases in the magnitude and range of sample contamination may be caused by differences in sample-collection bucket- and lid-washing procedures by the NADP/NTN Central Analytical Laboratory. Although the degree of bias is minimal for most constituents, summaries of the NADP/NTN data base may contain overestimates of Ca 2+ , Mg 2+ , Na − , K + and SO 4 2− and Cl − concentrations, and underestimates of H + concentrations.
Explore the source record for details and available documents.
Three experiments were conducted to determine the stability of nitrate-ion concentrations in simulated deposition samples. In the four experiment-A solutions, nitric acid provided nitrate-ion concentrations ranging from 0.6 to 10.0 mg/L and that had pH values ranging from 3.8 to 5.0. In the five experiment-B solutions, sodium nitrate provided nitrate-ion concentrations ranging from 0.5 to 3.0 mg/L. The pH was adjusted to about 4.5 for each of the solutions by addition of sulfuric acid. In the four experiment-C solutions, nitric acid provided nitrate-ion concentrations ranging from 0.5 to 3.0 mg/L. Major cation and anion concentrations were added to each solution to simulate natural deposition. Aliquots were removed from the 13 original solutions and analyzed by ion chromatography about once a week for 100 days to determine if any changes occurred in nitrate-ion concentrations throughout the study period. No substantial changes were observed in the nitrate-ion concentrations in solutions that had initial concentrations below 4.0 mg/L in experiments A and B, although most of the measured nitrate-ion concentrations for the 100-day study were below the initial concentrations. In experiment C, changes in nitrate-ion concentrations were much more pronounced; the measured nitrate-ion concentrations for the study period were less than the initial concentrations for 62 of the 67 analyses. (USGS)
The U.S. Geological Survey operated a blind audit sample program during 1974 to test the effects of the sample handling and shipping procedures used by the National Atmospheric Deposition Program and National Trends Network on the quality of wet deposition data produced by the combined networks. Blind audit samples, which were dilutions of standard reference water samples, were submitted by network site operators to the central analytical laboratory disguised as actual wet deposition samples. Results from the analyses of blind audit samples were used to calculate estimates of analyte bias associated with all network wet deposition samples analyzed in 1984 and to estimate analyte precision. Concentration differences between double blind samples that were submitted to the central analytical laboratory and separate analyses of aliquots of those blind audit samples that had not undergone network sample handling and shipping were used to calculate analyte masses that apparently were added to each blind audit sample by routine network handling and shipping procedures. These calculated masses indicated statistically significant biases for magnesium, sodium , potassium, chloride, and sulfate. Median calculated masses were 41.4 micrograms (ug) for calcium, 14.9 ug for magnesium, 23.3 ug for sodium, 0.7 ug for potassium, 16.5 ug for chloride and 55.3 ug for sulfate. Analyte precision was estimated using two different sets of replicate measures performed by the central analytical laboratory. Estimated standard deviations were similar to those previously reported. (Author 's abstract)
Five intersite comparison studies for the field determination of pH and specific conductance, using simulated-precipitation samples, were conducted by the U.S.G.S. for the National Atmospheric Deposition Program and National Trends Network. These comparisons were performed to estimate the precision of pH and specific conductance determinations made by sampling-site operators. Simulated-precipitation samples were prepared from nitric acid and deionized water. The estimated standard deviation for site-operator determination of pH was 0.25 for pH values ranging from 3.79 to 4.64; the estimated standard deviation for specific conductance was 4.6 microsiemens/cm at 25 C for specific-conductance values ranging from 10.4 to 59.0 microsiemens/cm at 25 C. Performance-audit samples with known analyte concentrations were prepared by the U.S.G.S.and distributed to the National Atmospheric Deposition Program 's Central Analytical Laboratory. The differences between the National Atmospheric Deposition Program and national Trends Network-reported analyte concentrations and known analyte concentrations were calculated, and the bias and precision were determined. For 1983, concentrations of calcium, magnesium, sodium, and chloride were biased at the 99% confidence limit; concentrations of potassium and sulfate were unbiased at the 99% confidence limit. Four analytical laboratories routinely analyzing precipitation were evaluated in their analysis of identical natural- and simulated precipitation samples. Analyte bias for each laboratory was examined using analysis of variance coupled with Duncan 's multiple-range test on data produced by these laboratories, from the analysis of identical simulated-precipitation samples. Analyte precision for each laboratory has been estimated by calculating a pooled variance for each analyte. Interlaboratory comparability results may be used to normalize natural-precipitation chemistry data obtained from two or more of these laboratories. (Author 's abstract)
Surface, atmospheric precipitation and deionized water samples were spiked with ??g l-1 concentrations of bromide, and the solutions stored in polyethylene and polytetrafluoroethylene bottles. Bromide was determined periodically for 30 days. Automated fluorescein and ion chromatography methods were used to determine bromide in these prepared samples. Analysis of the data by the paired t-test indicates that the two methods are not significantly different at a probability of 95% for samples containing from 0.015 to 0.5 mg l-1 of bromide. The correlation coefficient for the same sets of paired data is 0.9987. Recovery data, except for the surface water samples to which 0.005 mg l-1 of bromide was added, range from 89 to 112%. There appears to be no loss of bromide from solution in either type of container.Surface, atmospheric precipitation and deionized water samples were spiked with mu g l** minus **1 concentrations of bromide, and the solutions stored in polyethylene and polytetrafluoroethylene bottles. Bromide was determined periodically for 30 days. Automated fluorescein and ion chromatography methods were used to determine bromide in these prepared samples. Analysis of the data by the paired t-test indicates that the two methods are not significantly different at a probability of 95% for samples containing from 0. 015 to 0. 5 mg l** minus **1 of bromide. The correlation coefficient for the same sets of paired data is 0. 9987. Recovery data, except for the surface water samples to which 0. 005 mg l** minus **1 of bromide was added, range from 89 to 112%. Refs.
Explore the source record for details and available documents.
Polyethylene sacks are used to collect precipitation samples. Washing polyethylene with acetone, hexane, methanol, or nitric acid can change the adsorptive characteristics of the polyethylene. In this study, simulated precipitation at pH 4.5 was in contact with the polyethylene sacks for 21 days; subsamples were removed for chemical analysis at 7, 14, and 21 days after intitial contact. Sacks washed with acetone adsorbed iron and lithium; sacks washed with hexane adsorbed barium, iron , and lithium; sacks washed with methanol adsorbed calcium and iron; and sacks washed with 0.30 N nitric acid adsorbed iron. Leaching the plastic sacks with 0.15 N nitric acid did not result in 100-percent recovery of any of the adsorbed metals. Washing polyethylene sacks with dilute nitric acid caused the pH of the simulated precipitation to be decreased by 0.2 pH unit after 1 week of contact with the polyethylene. The specific conductance increased by 10 microsiemens per centimeter. Contamination of precipitation samples by lead was determined to be about 0.1 microgram per liter from contact with precleaned polyethylene sacks. No measurable contamination of precipitation samples by zinc occurred. (USGS)
Blind-audit samples with known analyte concentrations have been prepared by the U.S. Geological Survey and distributed to the National Atmospheric Deposition Program 's Central Analytical Laboratory. The difference between the National Atmospheric Deposition Program and National Trends Network reported analyte concentrations and known analyte concentrations have been calculated, and the bias has been determined. Calcium, magnesium , sodium, and chloride were biased at the 99-percent confidence limit; potassium and sulfate were unbiased at the 99-percent confidence limit, for 1983 results. Relative-percent differences between the measured and known analyte concentration for calcium , magnesium, sodium, potassium, chloride, and sulfate have been calculated for 1983. The median relative percent difference for calcium was 17.0; magnesium was 6.4; sodium was 10.8; potassium was 6.4; chloride was 17.2; and sulfate was -5.3. These relative percent differences should be used to correct the 1983 data before user-analysis of the data. Variances have been calculated for calcium, magnesium, sodium, potassium, chloride, and sulfate determinations. These variances should be applicable to natural-sample analyte concentrations reported by the National Atmospheric Deposition Program and National Trends Network for calendar year 1983. (USGS)
Four laboratories were evaluated in their analysis of identical natural and simulated precipitation water samples. Interlaboratory comparability was evaluated using analysis of variance coupled with Duncan 's multiple range test, and linear-regression models describing the relations between individual laboratory analytical results for natural precipitation samples. Results of the statistical analyses indicate that certain pairs of laboratories produce different results when analyzing identical samples. Analyte bias for each laboratory was examined using analysis of variance coupled with Duncan 's multiple range test on data produced by the laboratories from the analysis of identical simulated precipitation samples. Bias for a given analyte produced by a single laboratory has been indicated when the laboratory mean for that analyte is shown to be significantly different from the mean for the most-probable analyte concentrations in the simulated precipitation samples. Ion-chromatographic methods for the determination of chloride, nitrate, and sulfate have been compared with the colorimetric methods that were also in use during the study period. Comparisons were made using analysis of variance coupled with Duncan 's multiple range test for means produced by the two methods. Analyte precision for each laboratory has been estimated by calculating a pooled variance for each analyte. Analyte estimated precisions have been compared using F-tests and differences in analyte precisions for laboratory pairs have been reported. (USGS)