Potential sources of analytical bias and error in selected trace element data-quality analyses

Potential sources of analytical bias and error associated with laboratory analyses for selected trace elements where concentrations were greater in filtered samples than in paired unfiltered samples were evaluated by U.S. Geological Survey (USGS) Water Quality Specialists in collaboration with the USGS National Water Quality Laboratory (NWQL) and the Branch of Quality Systems (BQS). Causes for trace-element concentrations in filtered samples to exceed those in associated unfiltered samples have been attributed to variability in analytical measurements, analytical bias, sample contamination either in the field or laboratory, and (or) sample-matrix chemistry. These issues have not only been attributed to data generated by the USGS NWQL but have been observed in data generated by other laboratories. This study continues the evaluation of potential analytical bias and error resulting from matrix chemistry and instrument variability by evaluating the performance of seven selected trace elements in paired filtered and unfiltered surface-water and groundwater samples collected from 23 sampling sites of varying chemistries from six States, matrix spike recoveries, and standard reference materials. Filtered and unfiltered samples have been routinely analyzed on separate inductively coupled plasma-mass spectrometry instruments. Unfiltered samples are treated with hydrochloric acid (HCl) during an in-bottle digestion procedure; filtered samples are not routinely treated with HCl as part of the laboratory analytical procedure. To evaluate the influence of HCl on different sample matrices, an aliquot of the filtered samples was treated with HCl. The addition of HCl did little to differentiate the analytical results between filtered samples treated with HCl from those samples left untreated; however, there was a small, but noticeable, decrease in the number of instances where a particular trace-element concentration was greater in a filtered sample than in the associated unfiltered sample for all trace elements except selenium. Accounting for the small dilution effect (2 percent) from the addition of HCl, as required for the in-bottle digestion procedure for unfiltered samples, may be one step toward decreasing the number of instances where trace-element concentrations are greater in filtered samples than in paired unfiltered samples. The laboratory analyses of arsenic, cadmium, lead, and zinc did not appear to be influenced by instrument biases. These trace elements showed similar results on both instruments used to analyze filtered and unfiltered samples. The results for aluminum and molybdenum tended to be higher on the instrument designated to analyze unfiltered samples; the results for selenium tended to be lower. The matrices used to prepare calibration standards were different for the two instruments. The instrument designated for the analysis of unfiltered samples was calibrated using standards prepared in a nitric:hydrochloric acid (HNO 3 :HCl) matrix. The instrument designated for the analysis of filtered samples was calibrated using standards prepared in a matrix acidified only with HNO 3 . Matrix chemistry may have influenced the responses of aluminum, molybdenum, and selenium on the two instruments. The best analytical practice is to calibrate instruments using calibration standards prepared in matrices that reasonably match those of the samples being analyzed. Filtered and unfiltered samples were spiked over a range of trace-element concentrations from less than 1 to 58 times ambient concentrations. The greater the magnitude of the trace-element spike concentration relative to the ambient concentration, the greater the likelihood spike recoveries will be within data control guidelines (80–120 percent). Greater variability in spike recoveries occurred when trace elements were spiked at concentrations less than 10 times the ambient concentration. Spike recoveries that were considerably lower than 90 percent often were associated with spiked concentrations substantially lower than what was present in the ambient sample. Because the main purpose of spiking natural water samples with known quantities of a particular analyte is to assess possible matrix effects on analytical results, the results of this study stress the importance of spiking samples at concentrations that are reasonably close to what is expected but sufficiently high to exceed analytical variability. Generally, differences in spike recovery results between paired filtered and unfiltered samples were minimal when samples were analyzed on the same instrument. Analytical results for trace-element concentrations in ambient filtered and unfiltered samples greater than 10 and 40 μg/L, respectively, were within the data-quality objective for precision of ±25 percent. Ambient trace-element concentrations in filtered samples greater than the long-term method detection limits but less than 10 μg/L failed to meet the data-quality objective for precision for at least one trace element in about 54 percent of the samples. Similarly, trace-element concentrations in unfiltered samples greater than the long-term method detection limits but less than 40 μg/L failed to meet this data-quality objective for at least one trace-element analysis in about 58 percent of the samples. Although, aluminum and zinc were particularly problematic, limited re-analyses of filtered and unfiltered samples appeared to improve otherwise failed analytical precision. The evaluation of analytical bias using standard reference materials indicate a slight low bias for results for arsenic, cadmium, selenium, and zinc. Aluminum and molybdenum show signs of high bias. There was no observed bias, as determined using the standard reference materials, during the analysis of lead.

Scientific Investigations Report

Selenium in the upper Blackfoot River watershed, southeastern Idaho, 2001-12

The upper Blackfoot River in southeastern Idaho receives runoff from 12 large phosphate mines. Waste shales that are removed to access the phosphate ore are highly enriched with selenium, resulting in elevated selenium in runoff from the mine waste dumps. In 2001, in cooperation with the Bureau of Land Management, the U.S. Geological Survey (USGS) began monitoring streamflow, selenium, and other water-quality parameters at a single location near the outlet of the upper Blackfoot River to the Blackfoot Reservoir. Water samples primarily were collected by a flow triggered, automated pump sampler, supplemented by manual point and equal-width integrated manual samples. The approach to monitoring concentrations and streamflow over time at a fixed location is ideal for evaluating temporal trends, but provides no information about the relative source contributions from the mine waste dumps draining into various tributaries. In 2001, the Idaho Department of Environmental Quality (IDEQ) began an annual, mid-May, synoptic survey of selenium concentrations and streamflow at 21 locations along the main stem Blackfoot River and its tributaries. Individually, neither the intensive USGS sampling at the outlet nor the IDEQ annual synoptic sampling provides a comprehensive view of selenium runoff in the Blackfoot River watershed. Together, the efforts are complementary; therefore, in this report, results are presented from both sampling efforts. The USGS collected time-series data from 2001 to 2012 at a fixed location, the Blackfoot River near the outlet of the reservoir, near Henry, Idaho (USGS streamgage 13063000). Dissolved selenium concentrations from 450 filtered samples collected at this site ranged from 0.5 to 11.4 micrograms per liter (μg/L). The State of Idaho chronic aquatic life criterion concentration of 5 μg/L was exceeded in 31 percent of the samples, with most exceedances occurring during May of each year. No exceedances of the selenium criterion were recorded in months other than April, May, or June. Concentrations of selenium in unfiltered and filtered samples were similar, and concentrations from samples collected by depth and width integrated methods were similar to those collected by grab (point) samples, indicating that the grab samples adequately represent selenium concentrations across the entire river cross section. In speciation analyses made during 2003 and 2004, the median percentage of total selenium as selenate was 81 percent, ranging from 17 to 98 percent, and the median percentage of total selenium as selenite was 19 percent, ranging from 2 to 83 percent of the total selenium. During the period of study, selenium concentrations had an upward trend during the lowflow season of August–October. Time trends were not obvious during other seasons. Selenium daily loads varied by more than a factor of 900 during the study period and ranged from 0.03 kilograms per day (kg/d) to more than 24 kg/d. Annual maximum daily loads of selenium varied over nearly a factor of 12, ranging from about 2 to 24 kg/d. For the annual spring synoptic samples collected by the IDEQ along the main stem Blackfoot River and major tributaries, selenium concentrations ranged from less than 2 to 870 μg/L in 176 samples. In most years, the synoptic sampling showed that the majority of the selenium loads passing the USGS streamgage at the outlet of the watershed could be attributed to a single tributary, East Mill Creek, which enters the Blackfoot River through Spring Creek. Selenium loads decreased by about half from East Mill Creek before reaching the Blackfoot River, suggesting that much selenium is at least temporarily removed from the water column by uptake by aquatic vegetation or by losses to sediment. Similar decreases in selenium loads occurred through the main stem Blackfoot River before reaching the outlet in low flow years, but not in high flow years.

Idaho

Interpreting and Reporting Radiological Water-Quality Data

This document provides information to U.S. Geological Survey (USGS) Water Science Centers on interpreting and reporting radiological results for samples of environmental matrices, most notably water. The information provided is intended to be broadly useful throughout the United States, but it is recommended that scientists who work at sites containing radioactive hazardous wastes need to consult additional sources for more detailed information. The document is largely based on recognized national standards and guidance documents for radioanalytical sample processing, most notably the Multi-Agency Radiological Laboratory Analytical Protocols Manual (MARLAP), and on documents published by the U.S. Environmental Protection Agency and the American National Standards Institute. It does not include discussion of standard USGS practices including field quality-control sample analysis, interpretive report policies, and related issues, all of which shall always be included in any effort by the Water Science Centers. The use of 'shall' in this report signifies a policy requirement of the USGS Office of Water Quality.

Techniques and Methods

Chapter A5. Section 6.4.A. Arsenic Speciation

Two sample-processing methods (field speciation and laboratory speciation) used at the USGS National Water Quality Laboratory (NWQL) are specific to sample analysis by inductively coupled plasma-mass spectrometry (ICP-MS) for determining the concentration of inorganic and organic arsenic species in a water sample. The field-speciation method requires NWQL Schedule 1729. The laboratory-speciation method requires use either of NWQL Schedule 1730, 1731, or 1732, as appropriate for study objectives. For either the field- or laboratory-speciation method, prior knowledge is needed of sample matrix-composition characteristics (that is, major-ion concentrations in filtered samples). Major-ion data are necessary to determine (1) the volume of ethylenediaminetetraacetic acid (EDTA) that will be required for sample preservation, and (2) if sample dilution is required.

Techniques of Water-Resources Investigations

Mass spectrometric identification of an azobenzene derivative produced by smectite-catalyzed conversion of 3-amino-4-hydroxyphenylarsonic acid

The compound 3-amino-4-hydroxyphenylarsonic acid (3-amino-HPAA) reacts with smectite to form a soluble azobenzene arsonic acid compound. This reaction is of particular interest because it provides a possible mechanism for the formation of a new type of arsenic compound in natural water systems. 3-Amino-HPAA is a degradation product excreted by chickens that are fed rations amended with roxarsone. Roxarsone is used to control coccidial intestinal parasites in most of the broiler chickens grown in the United States. The structure of the azobenzene arsonic acid compound was first inferred from negative-ion and positive-ion low-resolution mass-spectrometric analyses of the supernatant of the smectite suspension. Elemental composition of the parent ion determined by high-resolution positive-ion mass spectrometric measurements was consistent with the proposed structure of the azobenzene arsonic acid compound.

Talanta

Inductively coupled plasma-mass spectrometry as an element-specific detector for field-flow fractionation particle separation

An inductively coupled plasma-mass spectrometer was used for the quantitative measurement of trace elements In specific,submicrometer size-fraction particulates, separated by sedimentation field-flow fractionation. Fractions were collected from the eluent of the field-flow fractionation centrifuge and nebulized, with a Babington-type pneumatic nebulizer, into an argon inductively coupled plasma-mass spectrometer. Measured Ion currents were used to quantify the major, minor, and trace element composition of the size-separated colloidal (< 1-microm diameter) particulates. The composition of surface-water suspended matter collected from the Yarra and Darling rivers in Australia is presented to illustrate the usefulness of this tool for characterizing environmental materials. An adsorption experiment was performed using cadmium lon to demonstrate the utility for studying the processes of trace metal-suspended sediment interactions and contaminant transport in natural aquatic systems.

Analytical Chemistry

Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory-Arsenic speciation in natural-water samples using laboratory and field methods

Analytical methods for the determination of arsenite [As(III)], arsenate [As(V)], dimethylarsinate (DMA), monomethylarsonate (MMA), and roxarsone in filtered natural-water samples are described. Various analytical methods can be used for the determination, depending on the arsenic species being determined. Arsenic concentration is determined by using inductively coupled plasma-mass spectrometry (ICP-MS) as an arsenic-specific detector for all methods. Laboratory-speciation methods are described that use an ion chromatographic column to separate the arsenic species; the column length, column packing, and mobile phase are dependent on the species of interest. Regardless of the separation technique, the arsenic species are introduced into plasma by eithe rpneumatic nebulization or arsine generation. Analysis times range from 2 to 8 minutes and method detection limits range from 0.1 to 0.6 microgram-arsenic per liter (ug-As/L), 10 to 60 picograms absolute (for a 100-microliter injection), depending on the arsenic species determined and the analytical method used. A field-generation specciation method also is described that uses a strong anion exchange cartridge to separate As(III) from As(V) in the field. As(III) in the eluate and the As(V) in the cartridge extract are determined by direct nebulization ICP-MS. Methylated arsenic species that also are retained on the cartridge will positively bias As(V) results without further laboratory separations. The method detection limit for field speciation is 0.3 ug-As/L. The distribution of arsenic species must be preserved in the field to eliminate changes caused by photochemical oxidation or metal oxyhydroxide precipitation. Preservation techniques, such as refrigeration, the addition of acides, or the additoin of ethylene-diaminetetraacetic acid (EDTA) and the effects of ambient light were tested. Of the preservatives evaluated, EDTA was found to work best with the laboratory- and field-speciation methods for all sample matrices tested. Storing the samples in opaque polytethylene bottles eliminated the effects of photochemical oxidation. The percentage change in As(III):As(V) ratios for an EDTA-preserved acid mine drainage (AMD) sample and ground-water sample during a 3-month period was -5 percent and +3 percent, respectively. The bias and variability of the methods were evaluated by comparing results for total arsenic and As(III), As(V), DMA, and MMA concentrations in ground water, AMD, and surface water. Seventy-one ground-water, 10 AMD, and 24 surface-water samples were analyzed. Concentrations in ground-water samples reached 720 ug-As/L for As(III) and 1080 ug-As/L for As(V); AMD samples reached 12800 ug-As/L for As(III) and 7050 ug-As/L for As(V); and surface-water samples reached 5 ug-As/L for As(III) and As(V). Inorganic arsenic species distribution in the samples ranged from 0 to 90 percent As(III). DMA and MMA were present only in surface-water samples from agricultural areas where the herbicide monosodium methylarsonate was applied; concentrations never exceeded 6 ug-As/L. Statistical analyses indicated that the difference between As(III) and As(V) concentrations for samples preserved with EDTA in opaque bottles and field-speciation results were analytically insignificant at the 95-percent confidence interval. There was no significant difference among the methods tested for total arsenic concentration. Percentage recovery for field samples spiked at 50 ug-As/L and analyzed by the laboratory-speciation method (n=2) ranged from 82 to 100 percent for As(III), 97 to 102 percent for As(V), 90 to 104 percent for DMA, and 81 to 96 percent for MMA; recoveries for samples spiked at 100 ug-As/L and analyzed by the field-speciation method ranged from 102 to 107 percent for As(III) and 105 to 106 percent for As(V). Laboratory-speciation results for Environment Canada reference material SLRS-2 closely matched reported concentrations. Laboratory-speciation metho

Water-Resources Investigations Report

Methods of analysis by the U. S. Geological Survey National Water Quality Laboratory-Determination of organic plus inorganic mercury in filtered and unfiltered natural water with cold vapor; atomic fluorescence spectrometry

An analytical method using cold vapor-atomic fluorescence spectrometry was developed by the U.S. Geological Survey in 2001 for the determination of organic plus inorganic mercury in filtered and unfiltered natural water. This method was developed to eliminate the use of acid dichromate preservative and to provide capability to measure ambient mercury concentrations in natural water. Dissolved mercury includes all oxidizable mercury species present in natural water that has been filtered through a 0.45- micrometer pore size capsule filter. Wholewater recoverable mercury includes dissolved mercury species and mercury species adsorbed to particulate matter in unfiltered natural water. Mercury species can include elemental mercury, mercury (II), mercury (II) complexes, various alkyl- and phenylmercury compounds, and other forms of mercury. In this method, samples are collected and processed according to standard U.S. Geological Survey protocols. Samples are preserved onsite with 6N hydrochloric acid in a ratio of 1 to 100 in a borosilicateglass bottle with fluoropolymer-lined cap. Mercury species are oxidized to mercury (II) by using bromine monochloride; excess oxidation reagent is neutralized with hydroxylamine hydrochloride. Elemental mercury produced after adding stannous chloride is purged from the solution with ultrapure argon gas into a cell in which the mercury concentration is measured by atomic fluorescence emission at 253.7 nanometers. The analytical response is linear up to 125 nanograms per liter (ng/L) of mercury, and the short-term method detection limit is about 5 ng/L. The analytical variability at 50 ng/L is about 10 percent. This report describes the method and compares the use of hydrochloric acid to acid dichromate as a field preservative. Ambient mercury concentrations in hydrochloric acidpreserved samples stored in borosilicate-glass bottles with fluoropolymer-lined caps are shown to be stable for at least 30 days. Mercury concentrations are stable for at least 5 months after bromine monochloride is added to the sample bottles in the laboratory. The long-term average percent recoveries at 20, 45, and 75 ng/L in reagent water, filtered and unfiltered ground water, and filtered and unfiltered surface water range from 89 to 108, 96 to 103, and 94 to 98 percent, respectively.

Water-Resources Investigations Report

Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory-Determination of whole-water recoverable arsenic, boron, and vanadium using inductively coupled plasma-mass spectrometry

Analysis of in-bottle digestate by using the inductively coupled plasma?mass spectrometric (ICP?MS) method has been expanded to include arsenic, boron, and vanadium. Whole-water samples are digested by using either the hydrochloric acid in-bottle digestion procedure or the nitric acid in-bottle digestion procedure. When the hydrochloric acid in-bottle digestion procedure is used, chloride must be removed from the digestate by subboiling evaporation before arsenic and vanadium can be accurately determined. Method detection limits for these elements are now 10 to 100 times lower than U.S. Geological Survey (USGS) methods using hydride generation? atomic absorption spectrophotometry (HG? AAS) and inductively coupled plasma? atomic emission spectrometry (ICP?AES), thus providing lower variability at ambient concentrations. The bias and variability of the methods were determined by using results from spike recoveries, standard reference materials, and validation samples. Spike recoveries in reagent-water, surface-water, ground-water, and whole-water recoverable matrices averaged 90 percent for seven replicates; spike recoveries were biased from 25 to 35 percent low for the ground-water matrix because of the abnormally high iron concentration. Results for reference material were within one standard deviation of the most probable value. There was no significant difference between the results from ICP?MS and HG?AAS or ICP?AES methods for the natural whole-water samples that were analyzed.

Open-File Report

Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory-Determination of arsenic and selenium in water and sediment by graphite furnace atomic absorption spectrometry

Graphite furnace-atomic absorption spectrometry (GF-AAS) is a sensitive, precise, and accurate technique that can be used to determine arsenic and selenium in samples of water and sediment. The GF-AAS method has been developed to replace the hydride generation-atomic absorption spectrometry (HG-AAS) methods because the method detection limits are similar, bias and variability are comparable, and interferences are minimal. Advantages of the GF-AAS method include shorter sample preparation time, increased sample throughput from simultaneous multielement analysis, reduced amount of chemical waste, reduced sample volume requirements, increased linear concentration range, and the use of a more accurate digestion procedure. The linear concentration range for arsenic and selenium is 1 to 50 micrograms per liter in solution; the current method detection limit for arsenic in solution is 0.9 microgram per liter; the method detection limit for selenium in solution is 1 microgram per liter. This report describes results that were obtained using stop-flow and low-flow conditions during atomization. The bias and variability of the simultaneous determination of arsenic and selenium by GF-AAS under both conditions are supported with results from standard reference materials--water and sediment, real water samples, and spike recovery measurements. Arsenic and selenium results for all Standard Reference Water Samples analyzed were within one standard deviation of the most probable values. Long-term spike recoveries at 6.25, 25.0, 37.5 micrograms per liter in reagent-, ground-, and surface-water samples for arsenic averaged 103 plus or minus 2 percent using low-flow conditions and 104 plus or minus 4 percent using stop-flow conditions. Corresponding recoveries for selenium were 98 plus or minus 13 percent using low-flow conditions and 87 plus or minus 24 percent using stop-flow conditions. Spike recoveries at 25 micrograms per liter in 120 water samples ranged from 97 to 99 percent for arsenic and from 82 to 93 percent for selenium, depending on the flow conditions used. Statistical analysis of dissolved and whole-water recoverable analytical results for the same set of water samples indicated that there is no significant difference between the GF-AAS and HG-AAS methods. Interferences related to various chemical constituents were also identified. Although sulfate and chloride in association with various cations might interfere with the determination of arsenic and selenium by GF-AAS, the use of a magnesium nitrate/palladium matrix modifier and low-flow argon during atomization helped to minimize such interferences. When using stabilized temperature platform furnace conditions where stop flow is used during atomization, the addition of hydrogen (5 percent volume/volume) to the argon minimized chemical interferences. Nevertheless, stop flow during atomization was found to be less effective than low flow in reducing interference effects.

Open-File Report

Organic compounds and trace elements in the Pocomoke River and tributaries, Maryland

In response to concern about recent blooms of the dinoflagellate, Pfiesteria piscicida, samples of sediment and water were collected from the lower Pocomoke River Basin and were screened for trace elements, pesticides, and other organic compounds. A large group of steroid and fatty acid methyl-ester compounds was detected in streamwater using gas chromatography/mass spectroscopy in scan mode. Some of these steroid compounds have been identified and further quantified in bed-sediment extracts. Spatial patterns of the concentrations of cholesterol suggest that these compounds are linked to the runoff of animal wastes into the river. Many of the organic compounds found in the Pocomoke River sediments have not yet been identified, but at least several are in the class of hormone compounds related to estradiols and have the potential to promote endocrine-disrupting effects in aquatic life. Particulate forms of arsenic and zinc are slightly elevated above normal levels for streams, but the sources for these elements are still undetermined. Several pesticides were found in low, parts-per-trillion concentrations, but were within the ranges commonly found in streams of this region.

Maryland

Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory-Determination of dissolved arsenic, boron, lithium, selenium, strontium, thallium, and vanadium using inductively coupled plasma-mass spectrometry

The inductively coupled plasma-mass spectrometric (ICP-MS) methods have been expanded to include the determination of dissolved arsenic, boron, lithium, selenium, strontium, thallium, and vanadium in filtered, acidified natural water. Method detection limits for these elements are now 10 to 200 times lower than by former U.S. Geological Survey (USGS) methods, thus providing lower variability at ambient concentrations. The bias and variability of the method was determined by using results from spike recoveries, standard reference materials, and validation samples. Spike recoveries at 5 to 10 times the method detection limit and 75 micrograms per liter in reagent-water, surface-water, and groundwater matrices averaged 93 percent for seven replicates, although selected elemental recoveries in a ground-water matrix with an extremely high iron sulfate concentration were negatively biased by 30 percent. Results for standard reference materials were within 1 standard deviation of the most probable value. Statistical analysis of the results from about 60 filtered, acidified natural-water samples indicated that there was no significant difference between ICP?MS and former USGS official methods of analysis.

Open-File Report

Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory-Comparison of a nitric acid in-bottle digestion procedure to other whole-water digestion procedures

A hydrochloric acid in-bottle digestion procedure is used to partially digest wholewater samples prior to determining recoverable elements by various analytical methods. The use of hydrochloric acid is problematic for some methods of analysis because of spectral interference. The inbottle digestion procedure has been modified to eliminate such interference by using nitric acid instead of hydrochloric acid in the digestion. Implications of this modification are evaluated by comparing results for a series of synthetic whole-water samples. Results are also compared with those obtained by using U.S. Environmental Protection Agency (1994) (USEPA) Method 200.2 total-recoverable digestion procedure. Percentage yields that use the nitric acid inbottle digestion procedure are within 10 percent of the hydrochloric acid in-bottle yields for 25 of the 26 elements determined in two of the three synthetic whole-water samples tested. Differences in percentage yields for the third synthetic whole-water sample were greater than 10 percent for 16 of the 26 elements determined. The USEPA method was the most rigorous for solubilizing elements from particulate matter in all three synthetic whole-water samples. Nevertheless, the variability in the percentage yield by using the USEPA digestion procedure was generally greater than the in-bottle digestion procedure, presumably because of the difficulty in controlling the digestion conditions accurately.

Open-File Report

Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory - Determination of elements in whole-water digests using inductively coupled plasma-optical emission spectrometry and inductively coupled plasma-mass spectrometry

Inductively coupled plasma-optical emission spectrometry (ICP-OES) and inductively coupled plasma-mass spectrometry (ICP-MS) can be used to determine 26 elements in whole-water digests. Both methods have distinct advantages and disadvantages--ICP-OES is capable of analyzing samples with higher elemental concentrations without dilution, however, ICP-MS is more sensitive and capable of determining much lower elemental concentrations. Both techniques gave accurate results for spike recoveries, digested standard reference-water samples, and whole-water digests. Average spike recoveries in whole-water digests were 100 plus/minus 10 percent, although recoveries for digests with high dissolved-solid concentrations were lower for selected elements by ICP-MS. Results for standard reference-water samples were generally within 1 standard deviation of hte most probable values. Statistical analysis of the results from 43 whole-water digest indicated that there was no significant difference among ICP-OES, ICP-MS, and former official methods of analysis for 24 of the 26 elements evaluated.

Open-File Report

Soluble trace elements and total mercury in Arctic Alaskan snow

Ultraclean field and laboratory procedures were used to examine trace element concentrations in northern Alaskan snow. Sixteen soluble trace elements and total mercury were determined in snow core samples representing the annual snowfall deposited during the 1993-94 season at two sites in the Prudhoe Bay oil field and nine sites in the Arctic National Wildlife Refuge (Arctic NWR). Results indicate there were two distinct point sources for trace elements in the Prudhoe Bay oil field- a source associated with oil and gas production and a source associated with municipal solid-waste incineration. Soluble trace element concentrations measured in snow from the Arctic NWR resembled concentrations of trace elements measured elsewhere in the Arctic using clean sample-collection and processing techniques and were consistent with deposition resulting from widespread arctic atmospheric contamination. With the exception of elements associated with sea salts, there were no orographic or east-west trends observed in the Arctic NWR data, nor were there any detectable influences from the Prudhoe Bay oil field, probably because of the predominant easterly and northeasterly winds on the North Slope of Alaska. However, regression analysis on latitude suggested significant south-to-north increases in selected trace element concentrations, many of which appear unrelated to the sea salt contribution.

Alaska