Determination of heat purgeable and ambient purgeable volatile organic compounds in water by gas chromatography/mass spectrometry

Two new analytical methods have been developed by the U.S. Geological Survey (USGS) National Water Quality Laboratory (NWQL) that allow the determination of 37 heat purgeable volatile organic compounds (VOCs) (USGS Method O-4437-16 [NWQL Laboratory Schedule (LS) 4437]) and 49 ambient purgeable VOCs (USGS Method O-4436-16 [NWQL LS 4436]) in unfiltered water. This report documents the procedures and initial performance of both methods. The compounds chosen for inclusion in the methods were determined as having high priority by the USGS National Water-Quality Assessment (NAWQA) Program. Both methods use a purge-and-trap technique with gas chromatography/mass spectrometry. The compounds are extracted from the sample by bubbling helium through a 25-milliliter sample. For the polar and less volatile compounds, the sample is heated at 60 degrees Celsius, whereas the less polar and more volatile compounds are purged using a separate analytical procedure at ambient temperature. The compounds are trapped on a sorbent trap, desorbed into a gas chromatograph/mass spectrometer for separation, and then identified and quantified. Sample preservation is recommended for both methods by adding a 1:1 solution of hydrochloric acid (HCl [1:1]) to water samples to adjust the pH to 2. Analysis within 14 days from sampling is recommended. The heat purgeable method (USGS Method O-4437-16) operates with the mass spectrometer in the simultaneous full scan/selected ion monitoring mode. This method supersedes USGS Method O-4024-03 (NWQL LS 4024). Method detection limits (MDLs) for fumigant compounds 1,2-dibromoethane, 1,2-dichloropropane, 1,2,3-trichloropropane, chloropicrin, and 1,2-dibromo-3-chloropropane range from 0.002 to 0.010 microgram per liter (µg/L). The MDLs for all remaining heat purgeable VOCs range from 0.006 µg/L for tert -butyl methyl ether to 3 µg/L for alpha -terpineol. Calculated holding times indicate that 36 of the 37 heat purgeable VOCs are stable for a minimum of 14 days preserved with HCl (1:1) to pH 2, and many are stable longer. Acrolein was retained in the method validation and initial method implementation and subsequently deleted because of instability and inconsistent performance. 2-Chloromethyl oxirane, methyl oxirane, and oxirane were tested using this method, but the compounds degraded quickly with the HCl (1:1) used for microbial preservation. The ambient purgeable method, USGS Method O-4436-16, operates with the mass spectrometer in the full scan mode. This method is a modification of USGS Method O-4127-96 (NWQL LS 2020). Several compounds were retained from Method O-4127-96 and will continue to be determined in Method O-4436-16. Eleven high priority compounds were added. MDLs for the high priority compounds range from 0.007 µg/L for 2,2-dichloro-1,1,1-trifluoroethane to 0.04 µg/L for 1,2,3,4-tetrahydronaphthalene and 1,3-butadiene. Historical MDLs for the compounds retained from Method O-4127-96 range from 0.009 µg/L for trans -1,2-dichloroethene to 0.1 µg/L for bromomethane. The calculated holding times for the compounds indicate the majority of the compounds are stable for a minimum of 14 days, or longer, at pH 2 with HCl (1:1) preservation. Four semivolatile compounds, 1,2-dimethylnaphthalene, 1,6-dimethylnaphthalene, 2,6-di- tert -butyl phenol, and 2-chloronapthalene, were tested and deleted from the method due to poor performance. Benzyl chloride was tested and deleted due to instability.

Techniques and Methods

Design and evaluation of a field study on the contamination of selected volatile organic compounds and wastewater-indicator compounds in blanks and groundwater samples

The Field Contamination Study (FCS) was designed to determine the field processes that tend to result in clean field blanks and to identify potential sources of contamination to blanks collected in the field from selected volatile organic compounds (VOCs) and wastewater-indicator compounds (WICs). The VOCs and WICs analyzed in the FCS were detected in blanks collected by the U.S. Geological Survey (USGS) National Water-Quality Assessment (NAWQA) Program during 1996–2008 and 2002–08, respectively. To minimize the number of variables, the study required ordering of supplies just before sampling, storage of supplies and equipment in clean areas, and use of adequate amounts of purge-and-trap volatile-grade methanol and volatile pesticide-grade blank water (VPBW) to clean sampling equipment and to collect field blanks. Blanks and groundwater samples were collected during 2008–09 at 16 sites, which were a mix of water-supply and monitoring wells, located in 9 States. Five different sample types were collected for the FCS at each site: (1) a source-solution blank collected at the USGS National Water Quality Laboratory (NWQL) using laboratory-purged VPBW, (2) source-solution blanks collected in the field using laboratory-purged VPBW, (3) source-solution blanks collected in the field using field-purged VPBW, (4) a field blank collected using field-purged VPBW, and (5) a groundwater sample collected from a well. The source-solution blank and field-blank analyses were used to identify, quantify, and document extrinsic contamination and to help determine the sources and causes of data-quality problems that can affect groundwater samples. Concentrations of compounds detected in FCS analyses were quantified and results were stored in the USGS National Water Information System database after meeting rigorous identification and quantification criteria. The study also utilized information provided by laboratory analysts about evidence indicating the presence of selected compounds, using less rigorous identification criteria than is required for reporting data to the National Water Information System database. For the FCS, these data are considered adequate to indicate "evidence of presence," and were used only for diagnostic purposes. Evidence of VOCs and WICs at low concentrations near or less than the long-term method detection level can indicate a contamination problem that could affect future datasets if method detection levels were ever to be lowered.

Scientific Investigations Report

Quality of volatile organic compound data from groundwater and surface water for the National Water-Quality Assessment Program, October 1996–December 2008

This report describes the quality of volatile organic compound (VOC) data collected from October 1996 to December 2008 from groundwater and surface-water sites for the U.S. Geological Survey's National Water-Quality Assessment (NAWQA) Program. The VOC data described were collected for three NAWQA site types: (1) domestic and public-supply wells, (2) monitoring wells, and (3) surface-water sites. Contamination bias, based on the 90-percent upper confidence limit (UCL) for the 90th percentile of concentrations in field blanks, was determined for VOC samples from the three site types. A way to express this bias is that there is 90-percent confidence that this amount of contamination would be exceeded in no more than 10 percent of all samples (including environmental samples) that were collected, processed, shipped, and analyzed in the same manner as the blank samples. This report also describes how important native water rinsing may be in decreasing carryover contamination, which could be affecting field blanks. The VOCs can be classified into four contamination categories on the basis of the 90-percent upper confidence limit (90-percent UCL) concentration distribution in field blanks. Contamination category 1 includes compounds that were not detected in any field blanks. Contamination category 2 includes VOCs that have a 90-percent UCL concentration distribution in field blanks that is about an order of magnitude lower than the concentration distribution of the environmental samples. Contamination category 3 includes VOCs that have a 90-percent UCL concentration distribution in field blanks that is within an order of magnitude of the distribution in environmental samples. Contamination category 4 includes VOCs that have a 90-percent UCL concentration distribution in field blanks that is at least an order of magnitude larger than the concentration distribution of the environmental samples. Fifty-four of the 87 VOCs analyzed in samples from domestic and public-supply wells were not detected in field blanks (contamination category 1), and 33 VOC were detected in field blanks. Ten of the 33 VOCs had a 90-percent UCL concentration distribution in field blanks that was at least an order of magnitude lower than the concentration distribution in environmental samples (contamination category 2). These 10 VOCs may have had some contamination bias associated with the environmental samples, but the potential contamination bias was negligible in comparison to the environmental data; therefore, the field blanks were assumed to be representative of the sources of contamination bias affecting the environmental samples for these 10 VOCs. Seven VOCs had a 90-percent UCL concentration distribution of the field blanks that was within an order of magnitude of the concentration distribution of the environmental samples (contamination category 3). Sixteen VOCs had a 90-percent UCL concentration distribution in the field blanks that was at least an order of magnitude greater than the concentration distribution of the environmental samples (contamination category 4). Field blanks for these 16 VOCs appear to be nonrepresentative of the sources of contamination bias affecting the environmental samples because of the larger concentration distributions (and sometimes higher frequency of detection) in field blanks than in environmental samples. Forty-three of the 87 VOCs analyzed in samples from monitoring wells were not detected in field blanks (contamination category 1), and 44 VOCs were detected in field blanks. Eight of the 44 VOCs had a 90-percent UCL concentration distribution in field blanks that was at least an order of magnitude lower than concentrations in environmental samples (contamination category 2). These eight VOCs may have had some contamination bias associated with the environmental samples, but the potential contamination bias was negligible in comparison to the environmental data; therefore, the field blanks were assumed to be representative. Seven VOCs had a 90-percent UCL concentration distribution in field blanks that was of the same order of magnitude as the concentration distribution of the environmental samples (contamination category 3). Twenty-nine VOCs had a 90-percent UCL concentration distribution in the field blanks that was an order of magnitude greater than the distribution of the environmental samples (contamination category 4). Field blanks for these 29 VOCs appear to be nonrepresentative of the sources of contamination bias to the environmental samples. Fifty-four of the 87 VOCs analyzed in surface-water samples were not detected in field blanks (category 1), and 33 VOC were detected in field blanks. Sixteen of the 33 VOCs had a 90-percent UCL concentration distribution in field blanks that was at least an order of magnitude lower than the concentration distribution in environmental samples (contamination category 2). These 16 VOCs may have had some contamination bias associated with the environmental samples, but the potential contamination bias was negligible in comparison to the environmental data; therefore, the field blanks were assumed to be representative. Ten VOCs had a 90-percent UCL concentration distribution in field blanks that was similar to the concentration distribution of environmental samples (contamination category 3). Seven VOCs had a 90-percent UCL concentration distribution in the field blanks that was greater than the concentration distribution in environmental samples (contamination category 4). Field-blank samples for these seven VOCs appear to be nonrepresentative of the sources of contamination bias to the environmental samples. The relation between the detection of a compound in field blanks and the detection in subsequent environmental samples appears to be minimal. The median minimum percent effectiveness of native water rinsing is about 79 percent for the 19 VOCs detected in more than 5 percent of field blanks from all three site types. The minimum percent effectiveness of native water rinsing (10 percent) was for toluene in surface-water samples, likely because of the large detection frequency of toluene in surface-water samples (about 79 percent) and in the associated field-blank samples (46.5 percent). The VOCs that were not detected in field blanks (contamination category 1) from the three site types can be considered free of contamination bias, and various interpretations for environmental samples, such as VOC detection frequency at multiple assessment levels and comparisons of concentrations to benchmarks, are not limited for these VOCs. A censoring level for making comparisons at different assessment levels among environmental samples could be applied to concentrations of 9 VOCs in samples from domestic and public-supply wells, 16 VOCs in samples from monitoring wells, and 9 VOCs in surface-water samples to account for potential low-level contamination bias associated with these selected VOCs. Bracketing the potential contamination by comparing the detection and concentration statistics with no censoring applied to the potential for contamination bias on the basis of the 90-percent UCL for the 90th-percentile concentrations in field blanks may be useful when comparisons to benchmarks are done in a study. The VOCs that were not detected in field blanks (contamination category 1) from the three site types can be considered free of contamination bias, and various interpretations for environmental samples, such as VOC detection frequency at multiple assessment levels and comparisons of concentrations to benchmarks, are not limited for these VOCs. A censoring level for making comparisons at different assessment levels among environmental samples could be applied to concentrations of 9 VOCs in samples from domestic and public-supply wells, 16 VOCs in samples from monitoring wells, and 9 VOCs in surface-water samples to account for potential low-level contamination bias associated with these selected VOCs. Bracketing the potential contamination by comparing the detection and concentration statistics with no censoring applied to the potential for contamination bias on the basis of the 90-percent UCL for the 90th-percentile concentrations in field blanks may be useful when comparisons to benchmarks are done in a study.

Scientific Investigations Report

Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory-Determination of gasoline oxygenates, selected degradates, and BTEX in water by heated purge and trap/gas chromatography/mass spectrometry

A method for determination of the alkyl ethers u sed as gasoline oxygenates [ethyl tert -butyl ether (ETBE), methyl tert -butyl ether (MTBE), diisopropyl ether (DIPE), and tert -amyl methyl ether (TAME)], some of their main degradates [acetone, methyl acetate, tert -butyl alcohol (tBA), and tert -amyl alcohol (tAA)], and BTEX (benzene, toluene, ethylbenzene, and xylenes) at low concentrations (<5 micrograms per liter) in water samples was developed. The compounds are determined using heated extraction to improve purging of polar compounds in a standard gas chromatography/ mass spectrometry (GC/MS) method for volatile compounds. Volatile compounds in this method are extracted (purged) from the sample b y bubbling helium through a 25-mL (milliliter) sample heated at about 65°C. Volatile compounds are trapped on a sorbent and then thermally desorbed into a GC/MS system for identification and quantitation. The calibration range for this method is 0.1 to 200 μg/L (micrograms per liter). Mean gasoline oxygenate recoveries from volatile-grade blank-water samples analyzed at concentrations from 0.5 to 5.0 μg/L were 95 to 105 percent, with relative standard deviations (RSDs) from 1.9 to 3.2 percent. Mean oxygenate degradate recoveries ranged from 88 to 107 percent, with RSDs of 3.2 to 7.4 percent, at concentrations from 1 to 50 μg/ L. Mean BTEX recoveries ranged from 91 to 107 percent, with RSDs of 1.1 to 6.6 percent, at concentrations from 0.5 to 10 μg/L . The method detection limits range from 0.035 to 0.052 μg/L for the gasoline oxygenates, 0.216 to 0.62 μg/L for the oxygenate degradates, and 0.005 to 0.036 μg/L for BTEX. Calculated holding times using American Society for Testing and Materials (ASTM) procedure D 4841-88 indicate that all of the analytes are stable for a minimum of 40 days at pH 2 and pH 7, except for methyl acetate, which is only stable for 7 days at pH 2.

Water-Resources Investigations Report

Laboratory method for analysis of small concentrations of methyl tert-butyl rther and other ether gasoline oxygenates in water

This Fact Sheet presents data for analysis of nanogram-per-liter concentrations of methyl tert-butyl ether (MTBE) and three other ether gasoline oxygenates, including methyl tert-pentyl ether (TAME), diisopropyl ether (DIPE), and ethyl tert-butyl ether (ETBE), by purge- and-trap capillary-column gas chromatography. Long-term method detection levels (LT-MDLs) for MTBE, TAME, DIPE, and ETBE ranged from 15 to 83 nanograms per liter (0.015 to 0.083 microgram per liter). Nanogram-per-liter-concentration detections are reported if all of the identification criteria are met, whereas previous methods censored detections at a pre-determined method reporting level. The reporting level for this method is defined as two times the LT-MDL, does not censor detections at less than this concentration, and is referred to as the nondetection value (NDV). Bias and variability data from multiple analyses, analysts, and instruments over a 60-day period show the oxygenate recoveries ranging from 100 to 109 percent, with 6 to 8 percent relative standard deviation. MTBE, TAME, DIPE, and ETBE were not detected in the analysis of 225 laboratory reagent blanks from January to December 1997. A preservation study in ground water and surface water indicates that all the oxygenates are stable at pH 2 for up to 216 days, with recoveries ranging from 94 to 115 percent on day 216, and relative standard deviations ranging from 5 to 9 percent for the duration of the study.

Fact Sheet

Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory-Determination of 86 volatile organic compounds in water by gas chromatography/mass spectrometry, including detections less than reporting limits

This report presents precision and accuracy data for volatile organic compounds (VOCs) in the nanogram-per-liter range, including aromatic hydrocarbons, reformulated fuel components, and halogenated hydrocarbons using purge and trap capillary-column gas chromatography/mass spectrometry. One-hundred-four VOCs were initially tested. Of these, 86 are suitable for determination by this method. Selected data are provided for the 18 VOCs that were not included. This method also allows for the reporting of semiquantitative results for tentatively identified VOCs not included in the list of method compounds. Method detection limits, method performance data, preservation study results, and blank results are presented. The authors describe a procedure for reporting low-concentration detections at less than the reporting limit. The nondetection value (NDV) is introduced as a statistically defined reporting limit designed to limit false positives and false negatives to less than 1 percent. Nondetections of method compounds are reported as ?less than NDV.? Positive detections measured at less than NDV are reported as estimated concentrations to alert the data user to decreased confidence in accurate quantitation. Instructions are provided for analysts to report data at less than the reporting limits. This method can support the use of either method reporting limits that censor detections at lower concentrations or the use of NDVs as reporting limits. The data-reporting strategy for providing analytical results at less than the reporting limit is a result of the increased need to identify the presence or absence of environmental contaminants in water samples at increasingly lower concentrations. Long-term method detection limits (LTMDLs) for 86 selected compounds range from 0.013 to 2.452 micrograms per liter (?g/L) and differ from standard method detection limits (MDLs) in that the LTMDLs include the long-term variance of multiple instruments, multiple operators, and multiple calibrations over a longer time. For these reasons, LTMDLs are expected to be slightly higher than standard MDLs. Recoveries for all of the VOCs tested ranged from 36 (tert-butyl formate) to 155 percent (pentachlorobenzene). The majority of the compounds ranged from 85 to 115 percent recovery and had less than 5 percent relative standard deviation for concentrations spiked between 1 to 500 ?g/L in volatile blank-, surface-, and ground-water samples. Recoveries of 60 set spikes at low concentrations ranged from 70 to 114 percent (1,2,3- trimethylbenzene and acetone). Recovery data were collected over 6 months with multiple instruments, operators, and calibrations. In this method, volatile organic compounds are extracted from a water sample by actively purging with helium. The VOCs are collected onto a sorbent trap, thermally desorbed, separated by a Megabore gas chromatographic capillary column, and finally determined by a full-scan quadrupole mass spectrometer. Compound identification is confirmed by the gas chromatographic retention time and by the resultant mass spectrum, typically identified by three unique ions. An unknown compound detected in a sample can be tentatively identified by comparing the unknown mass spectrum to reference spectra in the mass-spectra computer-data system library compiled by the National Institute of Standards and Technology.

Open-File Report